C.R. BARD, INC. (BASD) -3006260740 M.R.I. LOW-PROFILE IMPLANTABLE PORT, HICKMAN SINGLE-LUMEN, 6.6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
| Catalog Number |
0603880 |
| Medical Device Problem Codes |
Fracture (1260); Difficult to Insert (1316); Dull, Blunt (2407); Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a port placement procedure, when tried to dilate the vein, the internal dilator was allegedly noted with blunt tip.It was further reported that when tried to dilate, the internal dilator allegedly broke its tip.Furthermore, the device was removed.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, photos were provided for review.The investigation of the reported event is currently underway.H10: d4 (expiration date: 07/2026).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a port placement procedure, when tried to dilate the vein, the internal dilator was allegedly noted with blunt tip.It was further reported that when tried to dilate, the internal dilator allegedly broke its tip.Furthermore, the passage was not possible and the attempt to pass the catheter caused more bleeding than expected.Reportedly, the procedure was suspended and the device was removed.The procedure was completed by using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation.Three electronic photos were provided for review.The photo shows one the clinician showing the dilator.The distal tip of the dilator was noted to be blunt and deformed.Therefore the investigation is confirmed for the reported blunt and identified deformation issues.However the investigation is unconfirmed for the reported fracture issue as the reported event cannot be confirmed from the provided photos.Also the investigation is inconclusive for the reported difficulty in inserting issues as no sample was received to perform the functional tests.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: b5, d4 (expiration date: 07/2026), g3, h6 (device, method) h11: e1, h6 (result, conclusion) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a port placement procedure, when tried to dilate the vein, the internal dilator was allegedly noted with blunt tip.It was further reported that when tried to dilate, the internal dilator allegedly broke its tip.Furthermore, the passage was not possible and the attempt to pass the catheter caused more bleeding than expected.Reportedly, the procedure was suspended and the device was removed.The procedure was completed by using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one catheter, one bard port, one vessel dilator loaded into the distal end of a peel apart sheath and one peel apart sheath along with other components were received for evaluation.Functional, gross visual and microscopic visual evaluations were performed.The distal end of the peel-apart sheath was noted to be deformed and the distal end of the vessel dilator was noted to be bent; distal tip was noted to be deformed.The distal end of the peel-apart sheath was noted to be deformed as the edges appeared to be jagged.The bent distal tip of the vessel dilator was noted to be deformed.An attempt to load and retract the loaded vessel dilator from the peel-apart sheath was attempted and retracted without any issue.Both samples were noted to be bent.Also three electronic photos were provided for review.The photo shows one the clinician showing the dilator.The distal tip of the dilator was noted to be blunt and deformed.Therefore, the investigation is confirmed for the reported blunt and identified deformation issues.However, the investigation is unconfirmed for the reported fracture issue as the edges were noted to be jagged and deformed, and no fracture was noted.Also, the investigation is inconclusive for the reported difficulty in inserting issue as the exact circumstances at the time of the event was unknown.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 07/2026).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|