| Model Number |
H74939171040410 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon burst occurred.The 75% stenosed target lesion was located in the moderately tortuous and moderately calcified arteriovenous fistula.A 4mm x 40mmx 135cm mustang balloon catheter was advanced for dilation.During the procedure, the balloon burst upon first inflation at 10 atmospheres for 2 minutes.The device was removed successfully and there were no fragments left inside the patient.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
D2b: pro code (product code) - fge, lit.G4: premarket / 510(k) # - k103751, k110122.Device evaluated by manufacturer: returned product consisted of a mustang.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was observed inside the balloon which is evidence of a device leak.A leak test identified a balloon pinhole located at the distal edge of the distal markerband.The rated burst pressure for this device as per specification is 24 atmospheres.A visual and tactile examination found no kinks or damage to the shaft and tip of the device.A visual examination observed no issues with the markerbands of the device.Both markerbands were undamaged in the correct position on the device.No other issues were identified with the device.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon burst occurred.The 75% stenosed target lesion was located in the moderately tortuous and moderately calcified arteriovenous fistula.A 4mm x 40mmx 135cm mustang balloon catheter was advanced for dilation.During the procedure, the balloon burst upon first inflation at 10 atmospheres for 2 minutes.The device was removed successfully and there were no fragments left inside the patient.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Search Alerts/Recalls
|