• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD BACTISEAL CL EVD 1.9MM CTH SET; BACTISEAL EVD CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD BACTISEAL CL EVD 1.9MM CTH SET; BACTISEAL EVD CATHETERS Back to Search Results
Catalog Number 821750
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/05/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported a bactiseal catheter (id 821750) appeared to have broken in half when the patient moved while in the icu recovering from surgery.The catheter was implanted on (b)(6) 2024.A head computed tomography (ct) was performed to assure no parts left in the patient, but sufficient external catheter remained.The nurse noticed that the patient experienced a negative icp on monitor, a residual piece of catheter clamped and a presence of pneumocephalus on ct.The catheter was cut and reattached to collection system; however, the entire catheter was removed on (b)(6) 2024, and was not replaced.All broken parts were recovered.Patient status is progressing in care.
 
Additional Manufacturer Narrative
The bactiseal catheter was returned for evaluation.Failure analysis - the catheter was visually inspected; a tear was noted at the distal part.The catheter was irrigated, no occlusion was noted.The valve was leak tested and no leak noted.As only a part of device was returned the complaint could not be investigated more.Root cause analysis - the root cause for the issue reported by the customer, is due to improper handling of the device and could be caused by the patient moved.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BACTISEAL CL EVD 1.9MM CTH SET
Common Device Name
BACTISEAL EVD CATHETERS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18966520
Report Number3013886523-2024-00086
Device Sequence Number20475737
Product Code JXG
UDI-Device Identifier10886704040996
UDI-Public10886704040996
Combination Product (Y/N)N
PMA/510(K) Number
K090348
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number821750
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/24/2024
Supplement Report FDA Received Date04/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-