This device is classified as import for export, therefore 510k is not applicable.Model fb-15v is available in the usa with a 510k number k951199.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ocular fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ocular.In addition, our technician confirmed that the control body fluid damage, the forward body frame fluid damage, the biopsy inlet t-piece fluid damage, the cfb screen cover glass fluid damage, the cfb (coherent fiber bundle) fluid damage, and the insertion flexible tube perforated; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|