HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; FIBEROPTICCYSTOSCOPE 2.1C 4.9TP 400L ROP Back to Search Results
Model Number FCY-15RBS
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
The time of event is unknown.There was no report of patient harm.Fiber image failure(fluid damage).
 
Additional Manufacturer Narrative
This device is classified as import for export, therefore 510k is not applicable.Model fb-15v is available in the usa with a 510k number k951199.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ocular fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ocular.In addition, our technician confirmed that the control body fluid damage, the forward body frame fluid damage, the cfb screen cover glass fluid damage, the cfb (coherent fiber bundle) fluid damage, the insertion flexible tube coating damage, the insertion flexible tube perforated, the angle lever trim cap missing, the biopsy inlet barrel dirty, the biopsy inlet piece dirty, the biopsy inlet t-piece dirty, and the suction cylinder control body dirty; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
FIBEROPTICCYSTOSCOPE 2.1C 4.9TP 400L ROP
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
gurvinder nanda
3 paragon drive
montvale, NJ 07645
2015712318
MDR Report Key18966684
Report Number9610877-2024-52411
Device Sequence Number12435900
Product Code FAJ
UDI-Device Identifier04961333083664
UDI-Public04961333083664
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFCY-15RBS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2024
Initial Report FDA Received Date03/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-