| Catalog Number |
BOPS4311 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Osteolysis (2377)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the implant was removed due to bone loss & infection.Tooth location is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).A summary investigation has been completed for bone loss and infection events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss and infection have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost non-existent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
Upon follow up, implant placement and removal dates were updated.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to relay additional information.The following sections are being reported: b4: date of this report was updated.B5: event description was updated.D6: implant placement and removal dates were updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H10: narrative/data was updated.H3 other text : summary investigation.
|
| |
|
Event or Problem Description
|
|
No additional event information received at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.D9: device availability and return date were updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation code was added: 3331 and 4110.H10: narrative/data was updated.Dhr review was completed for the subject lot number 2022010181.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Sterilization record op#210 was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number 2022010181 for similar events and two other complaints were identified.As documented in the summary investigations, contributing factors for the reported event likely exist outside of zimvie control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigations, no malfunction occurred upon investigation.The reported events remain non-verifiable as they are a medical condition.Monthly trending to date has not identified any statistical triggers suggesting potential design or manufacturing related issues.Previously completed investigations for infection and/or bone loss, problems reported in association with implant failure have not identified any signals indicating potential non-conformances impacting the manufacturing and sterilization processes.Therefore, escalation to capa is not required.
|
| |
|
Search Alerts/Recalls
|