| Model Number |
N/A |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a total hip arthroplasty, the surgical nurse opened the implant to be delivered to the sterile field, but noticed the inside packaging was compromised and the plastic was crushed and damaged.There was no damage to the outer packaging.Surgery was completed using another device of the same size with no surgical delay.No additional information available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign: country: canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4;d9;g3;h2;h3;h6.Product was returned and evaluated.Visual review of the provided photos and returned product found that the sterile packaging blisters exhibit damage consistent with thermal damage.Sterility had been breached.Review of the device history records identified no deviations or anomalies during manufacturing.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.Medical records were not provided.The condition of the device when it left zimmer biomet is considered non-conforming to specification.The root cause of the reported event can be attributed to an operator error during packaging.This complaint was confirmed by evaluation of the provided photos and returned product.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|