Catalog Number ASK-45703-SL1
Medical Device Problem Codes Unraveled Material (1664); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/25/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported "threading wire through needle wouldn't pass, pulled wire out and discovered it was frayed".No medical intervention required.No patient harm or injury.The patient's current condition is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).The report that the guide wire was unravelled was confirmed through examination of the returned sample.The customer returned one guide wire for evaluation and the 18ga introducer needle was not returned.Visual examination revealed the guide wire was unravelled from the distal weld.The distal end of the core wire was broken and protruding out of the coil wire.Dimensional inspection and a device history record review based on a potential lot from sales history did not reveal any evidence of a manufacturing related issue.Functional inspection of the guide wire could not be performed for this complaint investigation due to the damage to the returned guide wire.The instructions-for-use (ifu) provided with this product warns the user, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained, and further consultation requested.Do not withdraw guidewire against needle bevel to reduce risk of possible severing or damaging of guidewire." arrow guide wires of this size are designed and manufactured to withstand a tensile force of 2.75 pounds force.This internal specif ication is higher than the bs en iso standard of 2.2 pounds force for this size wire.The selected insertion site and patient anatomy may present a tortuous path that could contribute to the possibility of guide wire kinking.Guide wire breakage may occur if a force greater than the design specification is applied during removal.Based on these circumstances, unintentional use error likely contributed to this event, however, the probable cause of guide wire and needle resistance could not be determined based upon the information provided and without the needle being returned for evaluation.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported "threading wire through needle wouldn't pass, pulled wire out and discovered it was frayed".No medical intervention required.No patient harm or injury.The patient's current condition is reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW MULTI LUMEN PI CVC KIT: 3L 7 FR
Common Device Name
CATHETER,INTRAVASCULAR,THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18968144
Report Number9680794-2024-00289
Device Sequence Number17717030
Product Code FOZ
UDI-Device Identifier10801902202366
UDI-Public10801902202366
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K071538
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberASK-45703-SL1
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2024
Supplement Date Received by Manufacturer03/28/2024
Initial Report FDA Received Date03/25/2024
Supplement Report FDA Received Date04/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED.; NONE REPORTED.