| Catalog Number |
RBY4C0860 |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/26/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.H3 other text : placeholder.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the superficial femoral artery using ruby coils and a microcatheter.During the procedure, the physician advanced the ruby coil into the target location and reported that the coil was not taking its intended shape.The physician made several attempts to get the ruby coil to take its intended shape without success.The physician then observed under fluoroscopy that the ruby coil had unintentionally detached inside the patient''s vessel.Therefore, the ruby coil was removed using forceps and a snare device.The procedure was completed using and another ruby coil, a non-penumbra coil, a vascular plug, and the same microcatheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Please note that the device associated with this complaint was expected to be returned.Further, additional information received from the distributor indicated that the device was sent out for return; however, an empty box was returned without the actual device.Therefore, without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|