• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number JC8386
Device Problems Material Twisted/Bent (2981); No Flow (2991); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/23/2024
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter last name: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an unspecified quantity of clearlink system continu-flo solution sets were kinked; the tubing was angled and flattened which caused a pump occlusion.This occurred during use of the devices, causing difficulty (flow issues) for the patients receiving their treatments.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information: d4: expiration date, h3, h4, h6, h10.D4: expiration date: october 2028.H10: the actual device was not available; however, photographs of a sample was provided for evaluation.Visual inspection of the photograph observed a kinked line which could generate an occlusion on the pump.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18968936
MDR Text Key338493581
Report Number1416980-2024-01305
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412152691
UDI-Public(01)00085412152691
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberJC8386
Device Lot NumberDR23I29079
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/27/2024
Initial Date FDA Received03/25/2024
Supplement Dates Manufacturer Received04/24/2024
Supplement Dates FDA Received04/25/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NI.; UNSPECIFIED PUMP.
-
-