| Catalog Number |
10014914 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4): initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Event or Problem Description
|
|
Material#: 10014914.It was reported by customer that leaking fluids.Upon infusion, rn noticed some leaking from gub connection between nad and pressure disc tubing.Verbatim#: rcc received a complaint via email.Email(s) attached please see the below cif / product complaint for item 10014914 on stocking po (b)(4).Please provide next steps on credit or return / credit.Record number : (b)(4).Initial reporter : (b)(6).Sales rep : (b)(6).Description as reported : leaking fluids.Upon infusion, rn noticed some leaking from gub connection between nad and pressure disc tubing.This is multiple items being reported as failed splitting cifs for each vendor - items are a pressure disc tubing, cardiac filter, microclave external reference number : (b)(4).Account number : (b)(6).Account name : (b)(6).Product code : al10014914.Additional information: what is the date of event? (b)(6) 2024.Describe any patient harm, injury, complication or negative outcome that occurred because of the event.The event occurred while in use with patient but no serious injury or harm was done.Delay in care.Any sample available for investigation? if yes, are you able to provide the address of the facility for us to ship the return label? yes, we have the item(s) ready to be returned for evaluation.(b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
The customer reported leaking at pressure disc, and returned one used sample of material 10014914, lot unknown.The set was visually examined for defects and abnormalities.No defects or abnormalities were observed.The set was infused with water from the included syringe, and the set infused normally without any leaks.The set was clamped and pressurized at 10 psi, but still, no leaks were observed.The customer complaint was not able to be replicated.The root cause is unknown without a replicated failure.A device history record review could not be performed on material 10014914 because the lot number is unknown.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
|
| |
|
Event or Problem Description
|
|
No additional information was provided.
|
| |
|