Catalog Number 159568
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Pain (1994)
Date of Event 02/29/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that ten years and six most post partial knee arthroplasty, the patient had to undergo a revision due to pain.During the procedure, it was noted that there was some wear to the bearing.Due diligence is in progress for this complaint; to date no additional information has been received.
 
Additional Manufacturer Narrative
(b)(4).D10 - associated medical devices: oxf uni tib tray sz b rm pma; item# 154721; lot# unknown.Oxf twin-peg cmntd fem sm pma; item# 161468; lot# unknown.Multiple mdr reports were filed for this event, please see associated reports: 3002806535-2024-00111 3002806535-2024-00112 investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.Multiple mdr reports were filed for this event, please see associated reports: 3002806535-2024-00111-2, 3002806535-2024-00112-2.Visual examination of the returned products identified signs of use (nicked, gouged) and confirming complaint has significant wear.Due to wear no measurements taken.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Review of the complaint histories identified additional similar complaints for the reported items and no additional similar complaints for the reported part and lot combinations.Radiographs were provided and reviewed by a radiologist.The review identified two views right knee dated over four (4) years after implantation demonstrates a medial compartment hemiarthroplasty with radiolucency along the cement hardware interface of the tibial component.Joint space narrowing could indicate polyethylene wear.Two views right knee dated about six (6) years after implantation demonstrates medial compartment hemiarthroplasty with persistent findings of polyethylene wear and possible loosening of the tibial component.Two views right knee dated eleven (11) years after implantation demonstrates medial compartment hemiarthroplasty with persistent findings of polyethylene wear and possible loosening of the tibial component.Medial compartment hemiarthroplasty of the right knee with possible polyethylene wear and loosening of the tibial component.Sizing is appropriate.Mild osteopenia.There is possible polyethylene wear and loosening of the tibial component.Possible polyethylene wear and loosening of the tibial component.Both of these findings can cause pain.With given information, a definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXFORD ANAT BRG RT SM SIZE 3 PMA
Common Device Name
KNEE PROSTHESIS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key18969545
Report Number3002806535-2024-00110
Device Sequence Number15005880
Product Code NRA
UDI-Device Identifier05019279786145
UDI-Public(01)05019279786145(17)170531(10)2682445
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P010014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/31/2017
Device Catalogue Number159568
Device Lot Number2682445
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/06/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/29/2024
Supplement Date Received by Manufacturer03/27/2024
07/17/2024
Initial Report FDA Received Date03/25/2024
Supplement Report FDA Received Date04/22/2024
07/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/04/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexPrefer Not To Disclose
Patient Weight70 KG
-
-