MAUDE Adverse Event Report: MICROVENTION, INC FRED X21; INTRALUMINAL DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC FRED X21; INTRALUMINAL DEVICES Back to Search Results
Model Number FREDX2518
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 03/04/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported fred x 21 was prepped per ifu.Delivered to the pericallosal artery.Successful delivery and implant.10 minutes post, the vessel was clotted off.Clot burden seen at the proximal end of the device.Integrilin was given to partially open the fredx 21.An atlas stent (4.0 x 24) was then delivered distal and proximal through the stent to ensure opening of the vessel.Patient was stable and doing fine post procedure on (b)(6) 2024, however reported back to the emergency room 15march2024 with general weakness.The general weakness was determined to have nothing to do with microvention device or the surgery.
 
Additional Manufacturer Narrative
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was implanted and therefore not available for return and investigation by the manufacturer.However, imaging was provided.The investigation is ongoing.Upon completion of the investigation, a supplemental mdr will be submitted.
 
Additional Manufacturer Narrative
Procedure/medical information review: imaging review, p24-1306: four radiographic images are supplied.None are identified as to date or time, or which vessel side is injected.P24-1306 -2 processed-85078276-169c-4312-a50f-8f633b0374d6: non-subtracted lateral head coned down x-ray, no contrast.A fred x 21 is seen in the a-2/3 segment of the pericallosal artery.It is well-opened.P24-1306 -3 processed-efabdf81-4f1f-4ed0-856d-7262f4290564: lateral coned-down ica subtracted dsa, with contrast.The fred x is again seen and there seems to be some faint clot in the proximal segment of the stent.There is an area of mild narrowing, in the a2 proximal to the stent, which is secondary to either vasospasm or atherosclerosis.There is a small, about 3 mm, pericallosal aneurysm covered by the fred.P24-1306 processed-11d4b6e5-7c50-4712-b1b4-c7732ed3e9fd: : lateral ica coned-down subtracted dsa, with contrast.The stent is now occluded by clot at its proximal segment.P24-1306 -4 processed-e60bd450-c731-4fc0-980c-c0332f08acd0: lateral ica coned-down subtracted dsa, with contrast.The stent is now open (presumably because of the integrilin) and two more stents (atlas, according to the complaint description) have been deployed.One, proximal and into the proximal part of the fred, and the second, spanning from the distal fred to just outside of it distally.The area of a2 spasm or atherosclerosis is again seen.There is no difference in aneurysm filling.These images do not directly explain why the fred clotted.However, this case is off-label (in the us) and the parent vessel diameter is small, which could have contributed.Investigation findings: items returned: n/a.Visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): potential complications below is a list of the probable adverse effects (e.G., complications) associated with the use of neurovascular flow diverting stents.Allergic reaction, including but not limited to: contrast dye, nitinol metal, and any other medications used during the procedure.Amaurosis fugax or transient blindness, aphasia, blindness, cardiac arrhythmia, complications of arterial puncture including pain, local bleeding, or injury to the artery, or adjacent nerves, cranial neuropathy, death, device fracture, migration or misplacement, diplopia, dissection or perforation of the parent artery, headache, hemiplegia, hemorrhage, including intracranial hemorrhage (ich), subarachnoid hemorrhage (sah), and retroperitoneal, hydrocephalus, infection, mass effect, myocardial infarction, neurological deficits, pseudoaneurysm formation, reactions to anti-platelet or anti-coagulant agents, reactions due to radiation exposure, including alopecia, burns ranging in severity from skin reddening to ulcers, cataracts, and delayed neoplasia.Reactions to anesthesia and related procedures, reactions to contrast agents including allergic reactions and kidney failure, reduced visual acuity or visual field, retinal artery occlusion or infarction, retinal ischemia, rupture or perforation of the aneurysm, stenosis of stented segment, seizure, stent thrombosis, stroke or tia (transient ischemic attack), thromboembolic event, vasospasm, visual impairment, warnings: the fred-27 system should only be delivered through a headway 27 microcatheter and the fred-21 system should only be delivered through a headway 21 microcatheter.If repeated friction is encountered during fred system delivery, verify microcatheter is not kinked or in extremely tortuous anatomy.Confirm that the microcatheter does not ovalize.Confirm that there is adequate sterile heparinized flush solution.Do not reposition the fred system in the parent vessel without fully retrieving the device.The fred system must be retrieved/re-sheathed into the microcatheter and re-deployed at the desired target location or removed completely from the patient.The fred system must not be re-deployed more than three times.Should unusual resistance be felt at any time during access or removal, the introducer/guide catheter/microcatheter and fred system should be removed as a single unit.Applying excessive force during delivery or retrieval of the fred system can potentially result in loss or damage to the device and delivery components.The fred system delivery wire should not be utilized as a guidewire.Do not torque the fred system.A torque device should not be used.Directions for use: 17.Advance the delivery wire to transfer the fred system from within the introducer into the microcatheter.18.Warning: do not torque the delivery wire while advancing or retracting the fred system.21.Warning: do not apply undue force.If resistance is encountered at any point during delivery or manipulation, withdraw the unit and select a new fred system.23.Position the fred system for deployment by aligning the fred system implant distal radiopaque end markers past the aneurysm neck, allowing for adequate distal and proximal device landing zones as shown in the following figures for fred-27 system and fred-21 system.26.Note: a slow, proper push/pull technique, encompassing sufficient delivery wire push force, in addition to an opposing microcatheter withdrawal force, to remove excess microcatheter slack while maintaining the microcatheter tip within the center of the parent vessel, will facilitate properly deploying the fred system at the proper location, to achieve full expansion and good vessel apposition.27.Caution: using a rapid microcatheter withdrawal technique to deploy the fred system is not recommended and may result in device elongation or improper deployment.Be aware of delivery wire tip position during deployment.28.If the fred system positioning is not satisfactory, the implant may be recaptured and repositioned if it is not fully deployed.The implant may be recaptured until the point where the distal-most wire marker, collocated distal to the implant proximal markers, is aligned approximately 50% of length proximal to the microcatheter distal marker band.29.Caution: if resistance is felt while recapturing the device, do not continue to recapture.Withdraw the microcatheter slightly to unsheath the device (without exceeding the recapture limit), and then attempt to recapture again.30.Caution: the fred system must not be re-deployed more than three times.31.If fred system positioning is satisfactory, carefully advance the delivery wire while retracting the microcatheter as needed to minimize slack, maintaining the microcatheter around the center of the parent vessel, to allow the implant to deploy across the neck of the aneurysm.Ensure the implant proximal radiopaque end markers are in the advised position (see step 23) proximal to the aneurysm neck for adequate coverage.32.Note: the fred system will expand and may foreshorten up to 60% from its undeployed length.Visually verify opening of the proximal end, ensuring that the microcatheter distal tip marker is pulled back, adequately away from the implant proximal end, to allow the proximal end to freely open.Push forward on the delivery wire to assist in maintaining access within the implant as needed.33.Note: visualize and refer to implant radiopaque end markers to maintain adequate implant length on each side of the aneurysm neck/target location to ensure appropriate coverage.34.Warning: do not fully deploy the fred system if positioning in the parent vessel is not satisfactory.37.Carefully inspect the deployed fred implant under fluoroscopy to confirm that it is completely open and opposed to the vessel wall and not kinked.If the implant is not fully open and apposed or is kinked, consider utilizing a suitable micro guidewire and/or occlusion balloon catheter to fully open the implant.38.Caution: carefully watch the fred implant distal and proximal markers when passing through the implanted device with other devices to avoid displacing the implant.Investigation conclusion: four radiographic images were provided for review.None of the images are identified as to date or time, or which vessel side is injected.The review of the images is as follows: p24-1306 -2 processed-85078276-169c-4312-a50f-8f633b0374d6: non-subtracted lateral head coned down x-ray, no contrast.A fred x 21 is seen in the a-2/3 segment of the pericallosal artery.It is well-opened.P24-1306 -3 processed-efabdf81-4f1f-4ed0-856d-7262f4290564: lateral coned-down ica subtracted dsa, with contrast.The fred x is again seen and there seems to be some faint clot in the proximal segment of the stent.There is an area of mild narrowing, in the a2 proximal to the stent, which is secondary to either vasospasm or atherosclerosis.There is a small, about 3 mm, pericallosal aneurysm covered by the fred.P24-1306 processed-11d4b6e5-7c50-4712-b1b4-c7732ed3e9fd: : lateral ica coned-down subtracted dsa, with contrast.The stent is now occluded by clot at its proximal segment.P24-1306 -4 processed-e60bd450-c731-4fc0-980c-c0332f08acd0: lateral ica coned-down subtracted dsa, with contrast.The stent is now open (presumably because of the integrilin) and two more stents (atlas, according to the complaint description) have been deployed.One, proximal and into the proximal part of the fred, and the second, spanning from the distal fred to just outside of it distally.The area of a2 spasm or atherosclerosis is again seen.There is no difference in aneurysm filling.These images do show the alleged clot in the proximal side of the fred stent as described in the reported event.The images do not directly explain why the fred clotted.However, this case is off-label (in the us) and the parent vessel diameter is small, which could have contributed to the issue.Without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.
 
Event or Problem Description
No additional information was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FRED X21
Common Device Name
INTRALUMINAL DEVICES
Manufacturer (Section D)
MICROVENTION, INC
35 enterprise
aliso viejo CA 92656
Manufacturer (Section G)
MICROVENTION, INC
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key18969684
Report Number2032493-2024-00242
Device Sequence Number6694073
Product Code OUT
UDI-Device Identifier00842429114223
UDI-Public(01)00842429114223(11)221108(17)251031(10)0000279857
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P180027
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Model NumberFREDX2518
Device Lot Number0000279857
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2024
Supplement Date Received by Manufacturer04/02/2024
Initial Report FDA Received Date03/25/2024
Supplement Report FDA Received Date04/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-