MEDOS INTERNATIONAL SARL HD COUPLER, C-MOUNT 17MM; ACCESSORIES, PHOTOGRAPHIC, FOR ENDOSCOPE (EXCLUDE LIGHT SOURCES)
|
Back to Search Results |
|
| Catalog Number |
242312 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the sales rep in hungary that during an unspecified surgical procedure on (b)(6) 2024 the hd coupler device c mount was foggy.Another coupler used to complete the case.There was no patient harm to the patient.There was a two minute delay while switching to another device.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device is not being returned, it was retained by the customer, therefore unavailable for a physical evaluation.Since the complaint device was not returned, we cannot determine a root cause for the reported failure.If additional information or the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.A manufacturing record evaluation was performed for the finished device, and no non-conformance's were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
| |
|
Search Alerts/Recalls
|
|
|