• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. MICPEG KIT WITH ENFIT CONNECTORS-20 FR-PULL; DH EF PEG INITIAL PLACEMENT PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. MICPEG KIT WITH ENFIT CONNECTORS-20 FR-PULL; DH EF PEG INITIAL PLACEMENT PRODUCTS Back to Search Results
Model Number 9640-20
Medical Device Problem Code Failure to Eject (4010)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 22-mar-2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(6).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.H3 other text : device not returned.
 
Event or Problem Description
It was reported the peg (percutaneous endoscopic gastrostomy) device "could not be traction removed in endo so [the device was] cut and retrieved" via endoscopy.The device was initially placed (b)(6)2023.
 
Additional Manufacturer Narrative
The device history record for the reported lot number, 30171964, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.All information reasonably known as of 19-jun-2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICPEG KIT WITH ENFIT CONNECTORS-20 FR-PULL
Common Device Name
DH EF PEG INITIAL PLACEMENT PRODUCTS
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT, INC.
6620 s. memorial place, suite 100
tucson AZ 85756
Manufacturer Contact
ujjal chakravartty
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key18970318
Report Number3006646024-2024-00010
Device Sequence Number12270904
Product Code KNT
UDI-Device Identifier00350770446431
UDI-Public(01)00350770446431(11)220201(17)231229(10)30171964
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K924065
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/29/2023
Device Model Number9640-20
Device Catalogue NumberN/A
Device Lot Number30171964
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2024
Supplement Date Received by Manufacturer06/02/2024
Initial Report FDA Received Date03/25/2024
Supplement Report FDA Received Date06/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-