| Model Number |
MAF-DM2 |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed that during the device inspection, the airway mobile scope exhibited connecting tube had coating peeling.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Correction to g3 of the initial medwatch.The aware date should be march 18, 2024.New information added to the following fields: h3 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over two (2) years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the defect was caused by stress of repeated use, external factors, or handling.The event can be prevented/detected by following the instructions for use described in the instruction manual operation manual: ¿chapter 3 preparation and inspection, 3.6 inspection of the endoscope¿.Olympus will continue to monitor field performance for this device.
|
| |
|