| Catalog Number |
381437 |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that bd insyte autoguard needle did not retract.The following information was provided by the initial reporter: needle safety did not work did the event involve an urgent/life threatening medical situation? no describe any patient harm, injury, complication or negative outcome that occurred as a result of the event.Please include treatment/intervention provided and the outcome.None reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Event or Problem Description
|
|
Did the event involve an urgent/life threatening medical situation? no.Describe any patient harm, injury, complication or negative outcome that occurred as a result of the event.Please include treatment/intervention provided and the outcome.None reported.Please confirm the number of occurrence.2.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: the complaint that the needle safety did not work was confirmed from the photograph that was provided for investigation.The photo showed an insyte autoguard needle and shield with evidence of use.The needle remained fully extended from the grip.Without the physical sample, the root cause of the damage could not be determined.A review of the inspection records and quality/manufacturing controls for the implicated lot indicated no issues with the manufacturing process.The complaint has been documented and will continue to be monitored as part of ongoing efforts to identify potential manufacturing related issues.
|
| |
|