| Model Number |
N/A |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device will not be returned for analysis as it remains implanted; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted h3 other text : device location is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon had broached for a size 5 taperloc however when he implanted it, it sat very proud and would not go down.The surgeon removed the stem and re-broached for a size 5 with the broach countersinking several mm.He then attempted a size 6 broach and it countersunk as well.Lastly, he reimplanted the stem for the third time and it seated within a few mm of complete seating.There is no additional information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected} updated: d4; g3; h2; h3; h4; h6.Proposed component code: mechanical (g04)- stem.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.The stem was not returned to confirm the size was conforming to specifications.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further information at the time of this report.
|
| |
|