COOK INC NEFF PERCUTANEOUS ACCESS SET; KGZ ACCESSORIES, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Codes Difficult to Remove (1528); Stretched (1601); Unraveled Material (1664)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
G4: pma/510(k) #: exempt.This report includes information known at this time.A follow-up report will be submitted should additional, relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported that the wire guide from a neff percutaneous access set unraveled during a percutaneous transhepatic biliary drainage procedure.After successful puncture with the access needle, bile flow was observed.Then the wire was introduced into the bile duct.After reaching the bile duct, it was found that the wire guide could not be withdrawn through the needle.Upon removal, it was discovered that the front end of the wire guide elongated and unraveled.Use of the wire was immediately discontinued.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unchanged, unknown, or unavailable.Correction: h6 (annex a).Investigation ¿ evaluation: it was reported that the wire guide from a neff percutaneous access set unraveled during a percutaneous transhepatic biliary drainage procedure.After successful puncture with the access needle, bile flow was observed.Then the wire was introduced into the bile duct.After reaching the bile duct, it was found that the wire guide could not be withdrawn through the needle.Upon removal, it was discovered that the front end of the wire guide elongated and unraveled.Use of the wire was immediately discontinued.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.Reviews of documentation including the complaint history, device history record (dhr), product labeling, quality control procedures, and specifications of the device were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The dhr for this complaint was reviewed.The final lot did not have any nonconformances.(b)(4).A complaint history search did not identify any other events associated with the reported device lot.Based on the available information, cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in the field.This product is not supplied with an instructions for use (ifu) pamphlet.The wire guide holder is supplied with an l_scor label attached indicating to not withdraw or manipulate the wire guide through the needle.Based on the information provided, no device return, and the results of the investigation, cook concludes this event to be due to failure to follow instructions.The customer withdrew the wire guide out of the needle which may have led to the device becoming stretched and elongated.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEFF PERCUTANEOUS ACCESS SET
Common Device Name
KGZ ACCESSORIES, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key18972222
Report Number1820334-2024-00416
Device Sequence Number20408070
Product Code KGZ
UDI-Device Identifier00827002105447
UDI-Public(01)00827002105447(17)260512(10)15437554
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/12/2026
Device Model NumberN/A
Device Catalogue NumberNPAS-100-RH-NT
Device Lot Number15437554
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/15/2024
Supplement Date Received by Manufacturer06/24/2024
Initial Report FDA Received Date03/25/2024
Supplement Report FDA Received Date07/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/12/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexFemale
-
-