• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER S.A.S. TORNIER FLEX SHOULDER SYS REV TRAY 6TH 1.5MM OFFSET; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. TORNIER FLEX SHOULDER SYS REV TRAY 6TH 1.5MM OFFSET; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number DWF511
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/27/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.H3 other text : device disposition unknown.
 
Event or Problem Description
It was reported that a patient required revision surgery due to infection.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.Microbiologist reviewed the sterilization procedures, environmental monitoring, bioburden data and the dhr and noted: the subject device was packaged according to established design and process specifications, and was sterilized according to procedure.No deviation for a non-conformance could be found.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Event or Problem Description
It was reported that a patient required revision surgery due to infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER FLEX SHOULDER SYS REV TRAY 6TH 1.5MM OFFSET
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR   38330
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18972587
Report Number3000931034-2024-00142
Device Sequence Number6571333
Product Code KWS
UDI-Device Identifier03700434019049
UDI-Public03700434019049
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K190521
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date04/27/2026
Device Catalogue NumberDWF511
Device Lot Number6711AW
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/28/2024
Supplement Date Received by Manufacturer05/08/2024
Initial Report FDA Received Date03/25/2024
Supplement Report FDA Received Date05/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-