BIOMET 3I CERTAIN LOW PROFILE ONE-PIECE ABUTMENT 3.4MM(D) X 2MM(H); DENTAL ABUTMENT Back to Search Results
Catalog Number ILPC342U
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/23/2023
Type of Reportable Event Malfunction
Event or Problem Description
The doctor reports that the dental abutment located in position number #14 failed because it fractured by the screw.The doctor reports that the procedure was not concluded by placing another abutment.Bone type: unknown.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).Zimvie received one (1) ilpc342u, (certain low profile one-piece abutment, 3.4mm(d) x 2mm(h) for evaluation.Visual evaluation was performed.Fracture identified at threads.Device history record (dhr) review was completed for the subject lot number 2022100421.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2022100421 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : functional : fracture : abutment.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was patient factors.Therefore, based on the available information, a device malfunction did occur.Fracture identified at threads.The reported event is confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN LOW PROFILE ONE-PIECE ABUTMENT 3.4MM(D) X 2MM(H)
Common Device Name
DENTAL ABUTMENT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key18973217
Report Number0001038806-2024-00560
Device Sequence Number6639157
Product Code NHA
UDI-Device Identifier00844868021100
UDI-Public(01)00844868021100(17)271021(10)2022100421
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K092341
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 08/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberILPC342U
Device Lot Number2022100421
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/25/2024
Supplement Report FDA Received Date08/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-