• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES EB210, VOYANT 5MM FUSION 37CM, 6/BX; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES EB210, VOYANT 5MM FUSION 37CM, 6/BX; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number EB210
Medical Device Problem Code Device Remains Activated (1525)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2024
Type of Reportable Event Malfunction
Event or Problem Description
Procedure performed: laparoscopic hysterectomy.Event description: long time voyant users, used device as intended.The device did not stop activating.Even when not holding the activation button.Intervention: device replacement patient status: no clinical impact to the patient.
 
Additional Manufacturer Narrative
The event unit is anticipated to return to applied medical for evaluation.A follow-up report will be submitted upon completion of investigation.
 
Event or Problem Description
Procedure performed: laparoscopic hysterectomy.Event description: long time voyant users, used device as intended.The device did not stop activating.Even when not holding the activation button.Additional information received by applied medical rep via email on 03apr24: the device continue to release rf energy after the button was released.The unintended output indeed occurred after releasing the button.The user stopped using the device after one time.Approx.15 times.It is unknown if there were alarms.The jaws were being cleaned as intended (long time voyant users).The number of times is not known.Intervention: device replacement.Patient status: no clinical impact to the patient.
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Testing was performed on the event unit.However, the complainant¿s experience could not be replicated or confirmed as the event unit passed all relevant testing and met current specifications.Applied medical has reviewed the details surrounding the event and related products and is unable to determine the cause of the reported event or confirm that a product malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EB210, VOYANT 5MM FUSION 37CM, 6/BX
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
farah azmi
22872 avenida empresa
rancho santa margarita, CA 92688
9497138710
MDR Report Key18974106
Report Number2027111-2024-00465
Device Sequence Number20386698
Product Code GEI
UDI-Device Identifier00607915138424
UDI-Public(01)00607915138424(17)260316(30)01(10)1486218
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K201212
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/16/2026
Device Model NumberEB210
Device Catalogue Number101476432
Device Lot Number1486218
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/25/2024
Supplement Report FDA Received Date07/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR; LAPAROSCOPIC INSTRUMENTS
Patient SexUnknown
-
-