| Model Number |
N/A |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2 - report source - foreign: event occurred in australia.The complainant has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.H3 other text : investigation incomplete.
|
| |
|
Event or Problem Description
|
|
It was reported that the tibial tray package was received damaged in transit and punctured through the plastic wrapping into the cardboard box.The contents of the packaging remained within the sterile plastic packaging.
|
| |
|
Additional Manufacturer Narrative
|
|
The product has been confirmed to exhibit outer packaging and/or labeling damage that did not result in a loss of sterility or label legibility.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.The reported event did not occur in an operating room or as part of a medical procedure.This complaint has been confirmed by review of the provided photo(s) and/or returned product.Based on the damage observed on the product, the damage to the outer packaging and/or labeling is consistent with damage from transit; therefore, the root cause of the reported event is attributed to transit damage.No corrective actions, preventive actions, or field actions resulted after the investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|