| Model Number |
VICM5_12.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A4: unk.A5: unk.A6: unk.H6 - work order search: no similar complaint was reported for units within the same lot.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 12.6mm vicm5_12.6 implantable collamer lens, -09.50 diopter, into the patient's right eye (od) on (b)(6) 2023.The lens was explanted on (b)(6) 2023 due to patient experienced a refractive surprise.The lens was replaced with another same model/length but different diopter lens and the problem was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation: the lens was returned dry, in microcentrifuge vial, with residue/ebris on product.Visual inspection found the haptic bent and deformed.Dimensional and refractive verification were performed and the lens was found to be in specification.(b)(4).
|
| |
|
Search Alerts/Recalls
|