• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MALLINCKRODT DAR SRL DAR; CATHETERS, SUCTION, TRACHEOBRONCHIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALLINCKRODT DAR SRL DAR; CATHETERS, SUCTION, TRACHEOBRONCHIAL Back to Search Results
Model Number 444SP02014
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code Low Oxygen Saturation (2477)
Date of Event 02/08/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, there was air leak in the closed suction system.The patient desaturated and had to be manually ventilated during troubleshooting and replacement of the suction system.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that there was air leak in the closed suction system.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAR
Common Device Name
CATHETERS, SUCTION, TRACHEOBRONCHIAL
Manufacturer (Section D)
MALLINCKRODT DAR SRL
via giacomo bove 2/4/6/8
mirandola 41037
IT  41037
Manufacturer (Section G)
MALLINCKRODT DAR SRL
via giacomo bove 2/4/6/8
mirandola 41037
IT   41037
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18975704
Report Number2936999-2024-00503
Device Sequence Number12337039
Product Code BSY
UDI-Device Identifier20884522035207
UDI-Public20884522035207
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K031997
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number444SP02014
Device Catalogue Number444SP02014
Device Lot Number22F0525FAX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer04/26/2024
Initial Report FDA Received Date03/25/2024
Supplement Report FDA Received Date04/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;