• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SALUDA MEDICAL PTY LTD EVOKE SCS SYSTEM; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SALUDA MEDICAL PTY LTD EVOKE SCS SYSTEM; SCS LEAD Back to Search Results
Model Number EVOKE CAP12 PERCUTANEOUS LEAD KIT - 60CM
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2024
Type of Reportable Event Serious Injury
Event or Problem Description
During the implant of an evoke spinal cord stimulation (scs) system, the lead was inserted into the closed loop stimulator (cls) header, the set screw was tightened and disconnected electrodes were identified.An attempt to reposition the lead in the cls was unsuccessful, as the lead could not be removed.The physician made the decision to complete the procedure with the disconnections since the second lead was showing impedances within range.A revision procedure was performed, and the lead was able to be removed and reinserted in the header.Following the procedure, the impedances were within range.
 
Additional Manufacturer Narrative
Investigation of this event is in progress.Once the investigation is complete, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Additional information: section g3 - date received by manufacturer.Section h6 - evaluation codes.Section h10 - manufacturer narrative.The evoke spinal cord stimulation (scs) system remains implanted.The cause of the reported disconnected electrode and lead removal difficulty could not be determined.Log review confirmed the reported event, and the reported disconnected electrodes were resolved post procedure.The manufacturing records were reviewed.The product met all requirements for release, and no anomalies were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOKE SCS SYSTEM
Common Device Name
SCS LEAD
Manufacturer (Section D)
SALUDA MEDICAL PTY LTD
5 eden park drive
macquarie park, nsw 2113
AS  2113
Manufacturer (Section G)
SALUDA MEDICAL PTY LTD
5 eden park drive
macquarie park, nsw 2113
AS   2113
Manufacturer Contact
sadie mauser
5 eden park drive
macquarie park, nsw 2113
AS   2113
MDR Report Key18976271
Report Number3021836309-2024-00215
Device Sequence Number12419501
Product Code LGW
UDI-Device Identifier09352307000773
UDI-Public09352307000773
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P190002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/16/2025
Device Model NumberEVOKE CAP12 PERCUTANEOUS LEAD KIT - 60CM
Device Catalogue Number3008
Device Lot Number9015604946
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2024
Supplement Date Received by Manufacturer04/20/2024
Initial Report FDA Received Date03/25/2024
Supplement Report FDA Received Date05/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-