| Catalog Number |
56538KA2/JPB |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/21/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The account alleges that during a procedure, the black tip of the catheter detached and remained within the blood vessel.The detached portion was successfully retrieved from the patient's blood vessel.No additional information available.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow-up report will be submitted once the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was returned for evaluation.The device was examined visually and microscopically.The complaint is confirmed.Root cause is attributed to a weak fuse joint.A review of the device history record was performed and no exception documents were identified.A review of the complaint history was performed and no similar complaints for this lot were found.Corrective actions are in process.
|
| |
|