BARD PERIPHERAL VASCULAR, INC. CONQUEST; PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number CQ5064
Medical Device Problem Codes Break (1069); Leak/Splash (1354); Difficult to Remove (1528); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an angioplasty procedure, when the pressure was pumped to 8 atm during repeat dilatation, pressure pump abnormality was noticed and blood was allegedly seen on gentle retraction.It was further reported that the pta balloon allegedly could not be fully retracted.Furthermore, upon removal of the balloon from the sheath, the balloon was allegedly found to be broken.The procedure was completed by using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a video was provided for review.The investigation of the reported event is currently underway.H10: d4.(expiration date: 10/2025).H11: section a: through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.One video was reviewed.The video shows the conquest balloon.Liquid is seen streaming from the proximal end of the balloon during an attempt to inflate it.Therefore, the investigation is confirmed for the reported balloon rupture as the balloon was seen leaking during inflation.However, the investigation is inconclusive for the reported removal difficulty as no objective evidence was provided.A definitive root cause for the reported balloon rupture and removal difficulty could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure, when the pressure was pumped to 8 atm during repeat dilatation, pressure pump abnormality was noticed and blood was allegedly seen on gentle retraction and the pta balloon allegedly could not be fully retracted.Reportedly, the balloon allegedly cracked and leaked.The procedure was completed by using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18976712
Report Number2020394-2024-00391
Device Sequence Number20408147
Product Code DQY
UDI-Device Identifier00801741062872
UDI-Public(01)00801741062872
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K083657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCQ5064
Device Lot NumberREGY0978
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/26/2024
Supplement Date Received by Manufacturer05/26/2024
Initial Report FDA Received Date03/26/2024
Supplement Report FDA Received Date06/01/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-