MAUDE Adverse Event Report: ABBOTT MEDICAL QUATTRODE LEAD WIDE SPACED, 30 CM; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL QUATTRODE LEAD WIDE SPACED, 30 CM; SCS LEAD Back to Search Results
Model Number 3163
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Codes Failure of Implant (1924); Inadequate Pain Relief (2388)
Date of Event 03/06/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Further information has been requested but not yet received.Date of event is estimated.
 
Event or Problem Description
Related manufacturer reference number: 1627487-2024-07730.It was reported that following multiple falls patient experienced ineffective therapy, both patients leads have high impedances and patient is unable to enter mri mode.As a result, surgical intervention may be undertaken to address the issue.
 
Event or Problem Description
Additional information indicates that patient was also experiencing pain at ipg site.Surgical intervention was undertaken on (b)(6) 2024 wherein ipg and leads were explanted and replaced to address the issue.
 
Additional Manufacturer Narrative
Correction e1- initial reporter information updated.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUATTRODE LEAD WIDE SPACED, 30 CM
Common Device Name
SCS LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key18977955
Report Number1627487-2024-07731
Device Sequence Number6224235
Product Code LGW
UDI-Device Identifier05414734406093
UDI-Public05414734406093
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/28/2016
Device Model Number3163
Device Catalogue Number3163
Device Lot Number4496441
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2024
Supplement Date Received by Manufacturer05/02/2024
06/03/2024
Initial Report FDA Received Date03/26/2024
Supplement Report FDA Received Date05/24/2024
06/11/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/03/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCS ANCHOR (X2); SCS IPG
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight98 KG
-
-