| Catalog Number |
03263991190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The cobas 8000 cobas c 502 module serial number was (b)(6).The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable creatinine (crep gen.2) results from the cobas 8000 cobas c 502 module.The patient was a transplant patient and was regularly checked by his nephrologist who claimed the results were too high and did not fit the patient history and status.The customer questioned if there was an interference with the patient's medications.The creatinine result using the jaffe method in another laboratory was <100 ¿mol/l.The creatinine results using the enzymatic method were 223 ¿mol/l and 177 umol/l.The results were obtained with different samples from different venipunctures.
|
| |
|
Additional Manufacturer Narrative
|
|
As no sample material was available, further investigation was not possible.The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|