|
MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507595MC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
03/06/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made. as such, the investigation will be closed. if the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803. this report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date. this report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report. reason for device explant and/or reoperation: size exchange.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a primary breast augmentation surgery with implantation of a 595cc mentor memorygel breast implant prosthesis.Post-operatively, the patient underwent a size exchange surgery to implant 385cc (left-sided) and 340cc (right-sided) mentor memorygel breast implants on (b)(6) 2024.During the surgery, bilaterally ruptured breast implants were discovered.There were no reported procedural delays or patient consequences due to the discovery.This report is for the right breast prosthesis.Refer to manufacturing report number 1645337-2024-03643 for the contralateral event.
|
| |
|
Additional Manufacturer Narrative
|
|
On march 26, 2024, the mentor analysis lab received the suspect medical device for evaluation.On april 01, 2024, mentor completed an evaluation on the returned device.Mentor conducted a visual inspection, microscopic examination, and thickness measurement of the device.Visual analysis of the returned sample revealed that the breast implant was found to be ruptured.A microscopic examination was performed, and the cause of the tear could not be identified.Therefore a thickness measurement was conducted on the shell at the tear, and the thickness was within manufacturing specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause the implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time. manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|