• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM512.6
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code Toxic Anterior Segment Syndrome (TASS) (4469)
Date of Event 02/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A4-a6:unk.H6: work order search: no similar complaints within associated lots were found.Manufacturers narrative: a review of the device labeling was completed.Iritis and cells in the anterior chamber are identified in the labeling as known potential adverse events following icl implantation.The dfu provides the surgeon instruction for complete ovd removal.Precaution: (6) after inserting this product, aspirate the viscoelastic substance completely from inside the eye.Do not use highly viscous viscoelastic substances that are difficult to aspirate completely.Toxic anterior segment syndrome (tass) is an acute sterile postoperative inflammation that can occur after uncomplicated or complicated intraocular surgery.While improper sterilization and contamination of surgical instruments remains the most common risk factor associated with tass, ocular viscoelastic, and improper preparation of antibiotics for intracameral injection may also lead to tass.An exact cause could not be determined as the event could be multifactorial in nature including patient and/or procedure related factors.Requests for additional information has not been forthcoming.Claim#(b)(4).
 
Event or Problem Description
The reporter indicated that a 12.6mm vticm512.6 implantable collamer lens of a -8.5/1.5/010 (sphere/cylinder/axis) was implanted into the patient's left eye (os) on (b)(6) 2024.Toxic anterior segment syndrome (tass) was observed on day 1.The patient reportedly had difficulty seeing through the haze and occasionally experienced pain behind the eye.There is no report of elevated iop.Information provided on (b)(6) 2024 states a bcva of 20/29 as the problem is not resolved.The lens remains implanted.In the reporter's opinion, the cause of the event is unknown.
 
Additional Manufacturer Narrative
H6: type of investigation 3331 - device history record review indicates that the product has been manufactured within the established process parameters and that there is no indication that the manufacturing and/or processing of the device contributed to the complaint issue.Claim# (b)(4).
 
Additional Manufacturer Narrative
B5: inflammatory recovery was observed eight days post-implantation.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
800 royal oaks dr. suite:210
monrovia, CA 91016
6263037902-2394
MDR Report Key18979579
Report Number2023826-2024-01083
Device Sequence Number17762095
Product Code QCB
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model NumberVTICM512.6
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2024
Supplement Date Received by Manufacturer06/04/2024
03/27/2026
Initial Report FDA Received Date03/26/2024
Supplement Report FDA Received Date07/19/2024
08/27/2026
Date Device Manufactured08/31/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: SFC-45 - LOT# UNK; FOAM TIP PLUNGER - LOT# UNK; INJECTOR MODEL: MSI-PF - LOT# UNK
Patient Age24 YR
Patient SexFemale
-
-