• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS LACTATE DEHYDROGENASE; LACTATE DEHYDROGENASE ASSAY USING NAD/NADH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS LACTATE DEHYDROGENASE; LACTATE DEHYDROGENASE ASSAY USING NAD/NADH Back to Search Results
Catalog Number 03004732122
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2024
Type of Reportable Event Malfunction
Event or Problem Description
There was an allegation of questionable lactate dehydrogenate (ldhi gen.2) results from the cobas 6000 c 501 analyzer.The customer was switching to a new lot of reagent and performed a patient comparison.The samples were repeated on another cobas 6000 c 501 using new reagent lot 747707.The customer did not know which result was correct.Patient 1 initial result was 377 u/l and the repeat result was 245 u/l.Patient 2 initial result was 260 u/l and the repeat result was 153 u/l.Patient 3 initial result was 390 u/l and the repeat result was 245 u/l.Patient 4 initial result was 262 u/l and the repeat result was 153 u/l.
 
Additional Manufacturer Narrative
The cobas 6000 c 501 analyzer serial number was (b)(6).The investigation is ongoing.
 
Additional Manufacturer Narrative
The field application specialist performed calibration and repeated the patient comparisons using reagent lots 747707 and 773098.The comparison was acceptable.The analyzer alarm trace contained multiple alarms indicating issues with sample quality.The investigation did not identify a product problem.The specific cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LACTATE DEHYDROGENASE
Common Device Name
LACTATE DEHYDROGENASE ASSAY USING NAD/NADH
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18979865
Report Number1823260-2024-00911
Device Sequence Number6570723
Product Code CFJ
UDI-Device Identifier04015630916276
UDI-Public04015630916276
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K964457
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Catalogue Number03004732122
Device Lot Number72604101
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/08/2024
Supplement Date Received by Manufacturer05/01/2024
Initial Report FDA Received Date03/26/2024
Supplement Report FDA Received Date05/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age15 YR
Patient SexFemale
-
-