| Model Number |
DRILL SAW SPORTS 400 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during an anterior cruciiate ligament btb surgery two pins broke off inside attachment head outside the patient.There was no harm for patient, operator or third party reported.The surgery was finished successfully with the same device anyway.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Event or Problem Description
|
|
Update dw 26-mar-2024: further information were provided that after disconnecting the attachment form hand piece, two metal pins fell out.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that during an anterior cruciate ligament btb surgery two pins broke off inside attachment head outside the patient.The reported event was confirmed.The manufacturer evaluation revealed to broken driver pins.A broken pin is most likely caused by placing too much load on the driver during use.
|
| |
|