| Model Number |
RETRACTABLE CANNULATED KNIFE, STRAIGHT |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 03/11/2024, it was reported by a sales representative via (b)(4) that (2) ar-6527-01 retractable cannulated knife handles could not slide even with pushing in the handle.This was discovered during a procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.Upon visual inspection, it was noted that the button of the retractable cannulated knife, straight had broken off.Functional testing was performed, and it was noted that the device did not work as required.The blade slid freely once and then got stuck.The most likely cause for the reported failure can be attributed to user error, due to the device due to excessive force/not being depressed when engaging the button to push forward.
|
| |
|