| Model Number |
39347-060470 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D2b pro code (product code): lit, dqy.Device evaluated by mfr.: the device was returned for analysis.As per the instructions for use (ifu), this device is compatible with a 0.035 inch (0.89 mm) guidewire.The guidewire used by the customer was not returned for analysis.The investigator successfully loaded the device on to a boston scientific 0.035 inch guidewire without issue.A visual examination of the returned device found that the balloon had been inflated.No issues were noted with the balloon material.A visual examination of the markerbands identified no issues.Both markerbands were present and undamaged in the correct position on the shaft.A visual and tactile examination found no kinks or damage to the shaft and tip of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon became stuck with wire.The 90% stenosed target lesions was located in the moderately tortuous and severely calcified vessel in the forearm.A 6.0mm x 40mm x 75cm athletis pta balloon dilatation catheter was advanced for dilation.During the procedure, the balloon became stuck with a 0.035 non-boston scientific guidewire.The guidewire was removed from the catheter outside the body.The procedure was completed with another of the same device.No complications were reported.
|
| |
|
Search Alerts/Recalls
|