• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II GUIDE WIRE Back to Search Results
Catalog Number 1009665
Medical Device Problem Codes Peeled/Delaminated (1454); Difficult to Advance (2920)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/05/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the ht bmw universal ii 300cm guidewire (gw) was attempted to be used in a procedure; however, the tip of the gw was catching during advancement on the introducer needle while the gw was still in the guide catheter.Therefore, the gw was removed from the patient and the teflon coating of the gw was noted to be missing and a transition bump was felt.It is unknown if anything was left in the anatomy.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported difficult to advance and the reported peeled / delaminated were unable to be confirmed.The coils were stretched for a length of 2mm.Electronic lot history record (lhr) and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint handling database identified no other incidents from this lot.As there was no damage noted to the guide wire during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that a coagulation of blood and/or contrast on the guide wire resulted in the reported difficult to advance.Manipulation of the device resulted in the noted stretched coils likely contributing to the reported difficult to advance.The reported peeled teflon coating was not confirmed.Based on the reported information and results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.E1 - facility name updated from (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005737652
road no.2 km 58.0 cruce dávila
barceloneta PR 00617
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18983812
Report Number2024168-2024-03724
Device Sequence Number2867835
Product Code DQX
UDI-Device Identifier08717648118722
UDI-Public(01)08717648118722(17)241231(10)3012771
Combination Product (Y/N)N
PMA/510(K) Number
K072460
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Catalogue Number1009665
Device Lot Number3012771
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer05/01/2024
Initial Report FDA Received Date03/26/2024
Supplement Report FDA Received Date05/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-