MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN POR CR HA FEM SZ 5 RT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN POR CR HA FEM SZ 5 RT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 71425215
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Loss of Range of Motion (2032)
Date of Event 02/27/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: internal complaint reference case (b)(4).
 
Event or Problem Description
It was reported that, after tka surgery had been performed on (b)(6) 2023, the patient experienced knee stiffness.This adverse event was solved by performing manipulation under anesthesia.Patient's current health status is recovered.No further complications were reported.
 
Additional Manufacturer Narrative
Section h3, h6: given the nature of the alleged incident, the devices could not be returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, per complaint details, the study patient experienced knee stiffness within 14 weeks of primary implantation.The case report form documented the mild severity adverse event onset date as (b)(6) 2024 as non-serious, not related to the study device but probably related to the procedure with a recovered/resolved outcome as of end date(b)(6) 2024 after concomitant procedure/surgery.The case report form show the patient mobilized post-op day 0 with 1 crutch as the primary ambulatory support at discharge and was prescribed post-op physical therapy/therapist-led home physical therapy.The case report form concomitant procedure sheet documented the procedure start and end date as (b)(6) 2024.Post-surgical adhesions/scar tissue and/or arthrofibrosis are known potential post-surgical complications and would not support a component malperformance.The impact to the patient beyond the reported knee stiffness and subsequent manipulation under anesthesia cannot not be determined based on the information provided.A review of the production orders did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.For the femoral component, a review of complaint history based on the historical data revealed similar events for the listed batch.For the patellar component and tibial baseplate, the review revealed similar events for the listed devices over the previous 12 months, but no similar events for the batches based on the historical data.For the insert, no similar events were revealed for the part number over the past 12 months nor for the batch number based on historical data.This failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed in possible adverse effects that decreased range of motion and unusual stress concentrations can result from trauma, improper implant selection, improper implant positioning, improper fixation, and/or migration of the components.Muscle and fibrous tissue laxity can also contribute to these conditions.A review of the risk management files revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this products and event.At this time, we have no evidence to conclude that the products failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include alignment, size selected or wear.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.H6: health effect - impact code.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN POR CR HA FEM SZ 5 RT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18983947
Report Number1020279-2024-00622
Device Sequence Number12828323
Product Code MBH
UDI-Device Identifier00885556108505
UDI-Public00885556108505
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K091543
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71425215
Device Lot Number20LHA0031
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer04/22/2024
Initial Report FDA Received Date03/26/2024
Supplement Report FDA Received Date04/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LGN XLP INS SZ 3-4 9MM(PN:71425338,LN:21DM13440); PRS RND PAT SZ 32MM(PN:74025464,LN:22GM06622A); PRS TIB BSPT SZ3 RT(PN:71425303,LN:23GM04886)
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age69 YR
Patient SexFemale
Patient Weight60 KG
-
-