STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5_12.6
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Endophthalmitis (1835)
Date of Event 02/21/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H6 - type of investigation - lens work order search: no similar complaint type events reported for units within the same lot.Manufacturer narrative: a review of the device labeling was completed.Endophthalmitis is identified in the labeling as a potential adverse event following icl implantation.Per the directions for use complications and adverse reactions implantation of any evo|evo+ icl may include but are not limited to: intraocular infection and secondary surgical intervention to remove/replace the lens.The dfu states a caution: staar surgical evo|evo+ ticl and disposable accessories are packaged and sterilized for single use only.Cleaning, reuse and/or resterilization are not applicable to these devices.If one of these devices were reused after cleaning and/or resterilization, it is highly probable that it would be contaminated, and the contamination could result in infection and/or inflammation.An exact cause could not be determined as the event could be multifactorial in nature including patient and/or procedure related factors.The lens was explanted and returned.Claim# (b)(4).
 
Event or Problem Description
The reporter indicated that a 12.6mm vticm5 implantable collamer lens of -10.50/+1.50/100 power was implanted into the patients left eye (os) on (b)(6) 2024.Postop ciproxacino and maxidex drops were given.On (b)(6) 2024 the lens was explanted due to observation of endophthalmitis.Anterior chamber wash and intravitreal corticosteroid/antibiotic treatment occurred.Oral antibiotics were also given.Diagnostic cultures were performed with lab results negative for organisms.There is no report of loss of visual acuity.On (b)(6) 2024 it was reported the patient has a suture in place and was doing well.The patient will undergo icl reimplantation two months later.
 
Additional Manufacturer Narrative
B5: per updated information the patient was seen, the stitches were removed and visual acuity is 20/20 with correction.The endophthalmitis was reported to be resolved.H6: inspection of the returned device identified four particulates on the lens.External laboratory analysis was performed to determine the size and composition of the particulates identified on the lens.First particle: found to be primarily composed of cellulosic material, second particle: primarily composed of protein and fat, third particle: primarily composed of amorphous carbon, with trace elements including oxygen, sodium, chlorine, nitrogen, aluminum, sulfur, and iron, fourth particle: primarily composed of stainless steel, with trace amounts of carbon, oxygen, silicon, sodium, chlorine, aluminum, and nitrogen.Unfortunately, we were unable to conclusively identify the source of the elements found in the sem-edx and ftir analyses.H6- device history record (dhr) review: based on the results of the investigation, the dhr review indicates that the product has been manufactured within the established process parameters and there is no indication that the manufacturing and/or processing of the device contributed to the complaint issue.Claim # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key18984029
Report Number2023826-2024-01347
Device Sequence Number15555001
Product Code QCB
UDI-Device Identifier00840311308118
UDI-Public00840311308118
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberVTICM5_12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer05/24/2024
Initial Report FDA Received Date03/26/2024
Supplement Report FDA Received Date08/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: SFC-45 W/FTP, LOT# UNK; INJECTOR MODEL: MSI-PF, LOT# UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age28 YR
Patient SexFemale
-
-