Catalog Number RED72KIT
Medical Device Problem Codes Fracture (1260); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/26/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure to treat a large vessel occlusion (lvo) in the left m1 segment of the middle cerebral artery (mca) using a penumbra system red72 reperfusion catheter (red72), a neuron max 6f 088 long sheath (neuron max), a non-penumbra stent retriever and a guidewire.During the procedure, the physician deployed the stent retriever and completed one pass using the red72.After completion of the first pass, the physician retracted the stent retriever through the red72.While retracting the red72, the physician experienced resistance and noticed under fluoroscopy that the red72 was fractured at the distal length.The physician then attempted to aspirate the fractured distal segment of the red72 using a new red72 but was unable to retrieve it.The procedure ended at this point.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM RED 72 REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key18984188
Report Number3005168196-2024-00109
Device Sequence Number8204253
Product Code NRY
UDI-Device Identifier00815948023932
UDI-Public815948023932
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
K211654
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/26/2024
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/17/2026
Device Catalogue NumberRED72KIT
Device Lot NumberH00003236
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2024
Initial Report FDA Received Date03/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexMale
Date Report Sent to FDA01/01/2005
-
-