| Model Number |
RT019 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) fisher & paykel healthcare has received the subject device for evaluation.We will provide a follow up report upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
A distributor reported on behalf of a healthcare facility in japan that a rt380 adult dual heated evaqua2 breathing circuit failed the pre-use ventilator leak test.It was also reported that the source of the leak was likely to be the rt019 inspiratory/expiratory filter provided as part of the rt380 adult dual heated evaqua2 breathing circuit.There was no patient involvment.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Method: the complaint (b)(4).Inspiratory/expiratory filter was returned to fisher & paykel healthcare (f&p) in new zealand for investigation, where it was visually inspected.Results: visual inspection identified that the filter housing had been misaligned during the manufacturing process.Conclusion: the filter was incorrectly assembled during production.An incorrectly assembled filter would be detected prior patient use, during the pre-use leak testing, as reported in this event.The user instructions supplied with the rt380 breathing circuit state: - visually inspect breathing sets for damage (e.G.A crushed tube or cracked connector) before use, and replace if damaged.- perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.- set appropriate ventilator alarms.
|
| |
|
Event or Problem Description
|
|
A distributor reported on behalf of a healthcare facility in japan that a rt380 adult dual heated evaqua2 breathing circuit failed the pre-use ventilator leak test.It was also reported that the source of the leak was likely to be the rt019 inspiratory/expiratory filter provided as part of the rt380 adult dual heated evaqua2 breathing circuit.There was no patient involvment.
|
| |
|