• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APIFIX LTD. MID-C 125; POSTERIOR RATCHETING ROD SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APIFIX LTD. MID-C 125; POSTERIOR RATCHETING ROD SYSTEM Back to Search Results
Model Number MID-C 115
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2024 apifix was notified that patient #446 (index procedure (b)(6) 2020) is scheduled for revision surgery of (b)(6) 2024 due to a broken rod (male portion of the ratchet mechanism).Minimal symptoms.
 
Additional Manufacturer Narrative
Production process analysis: a review of the device history record confirmed that the device was manufactured and tested according to relevant procedures, and shipped according to manufacturer's specifications.User (surgeon and patient) information analysis: on 29-feb-2024 apifix was notified that patient #(b)(6) (index procedure 04-dec-2020) is scheduled for revision surgery of 05-apr-2024 due to a broken rod (male portion of the ratchet mechanism).Minimal symptoms.Risk assessment: implant breakage can result from trauma, practicing contact or high demand sports, hyper-kyphosis, inserting the pedicle screws in a wrong trajectory which locks the poly-axial joints, not working according to the surgical technique (e.G., minimally invasive approach and/or not using the trial tool properly).Re-operation events are a known risk that was assessed and recorded by the product risk assessment.The risk of mechanical failure of the mid-c mechanism - [implant breakage] has been assessed and found to be acceptable.The implant breakage rate due to any reason is in line with the rate reported in the literature for this type of complication as described in the company's clinical evaluation report.The explanted device is expected to be returned to the manufacturer for analysis after the revision surgery.Upon completion of the evaluations, when additional information comes to light, then a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
On (b)(6) 2024 the patient underwent revision surgery during which the mid-c was replaced with a new one, and new polyaxial screws were implanted at t5-t6.The apifix screw in t12 remained implanted and used with the construct.Return analysis: the explanted device was returned, and was subjected to cleaning, steam sterilizing, and engineering evaluation.The device was obviously fractured in the pole component around the mid-point of the pole.Post-fracture, the device remained implanted as the fracture plane, inside of the body, and edge of the base and pole showed signs of wear.The fracture plane was worn smooth and thus fracture mode is unable to be identified.The extender was removed, and the spherical rings were inspected for wear.Wear and deformity were observed on the spherical ring of the base.No wear was visibly observed on the spherical ring of the pole.Over the years, the company already implemented corrective action with the following: on (b)(6) 2017, all surgeons received a letter detailing the importance of screw insertion trajectory.The topic is also covered in the company's training presentation.Eco-38 (approved in h170001/s002): replaced the mid-c 125 that extends by 40 mm to mid-c 125 that extends by 50mm allowing more overlap between the pole and base.Eco-46 (approved in h170001/s002): a trial tool added to the surgical tools to aid the surgeon in detecting if access tissue remains below the implant.On march 2020, the topic of practicing severe sports added to the mid-c training presentation.A precaution about severe sports was included in the ifu approved in h170001.On october 2020, as part of apifix commitment to continuous improvement, capa #020-01 was initiated to further investigate to prevent and minimize the rate of implant breakage.Eco-13264 (approved in h170001/s015): increase strength through: decrease stress concentrations and increase cross sectional area at critical regions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MID-C 125
Common Device Name
POSTERIOR RATCHETING ROD SYSTEM
Manufacturer (Section D)
APIFIX LTD.
1 hacarmel st.
yokneam ellit, 20692 07
IS  2069207
Manufacturer (Section G)
APIFIX LTD.
1 hacarmel st.
kochav yokneam bldg
yokneam ellit, 20692 07
IS   2069207
Manufacturer Contact
alan vaisman
1 hacarmel st.
kochav yokneam bldg
yokneam ellit, 20692-07
IS   2069207
MDR Report Key18984849
Report Number3013461531-2024-00015
Device Sequence Number2858755
Product Code QGP
UDI-Device Identifier07290018128046
UDI-Public07290018128046
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
H170001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMID-C 115
Device Catalogue NumberMUS-115-040
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/29/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/27/2024
Supplement Report FDA Received Date07/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age16 YR
Patient SexMale
-
-