• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. CLARIA MRI CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. CLARIA MRI CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTMA1D4
Medical Device Problem Codes Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 03/04/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the patient received inappropriate anti-tachycardia pacing and inappropriate high voltage therapy for supra ventricular tachycardia (svt) that the cardiac resynchronization therapy defibrillator (crt-d) detected as ventricular tachycardia (vt).It was noted that the therapies were initially appropriately withheld.The crt-d remains in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARIA MRI CRT-D SURESCAN
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18986137
Report Number3004209178-2024-07891
Device Sequence Number2894949
Product Code NIK
UDI-Device Identifier00643169929548
UDI-Public00643169929548
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P010031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/14/2019
Device Model NumberDTMA1D4
Device Catalogue NumberDTMA1D4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2024
Initial Report FDA Received Date03/27/2024
Date Device Manufactured10/17/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
419688 LEAD, 6947M55 LEAD, 5076-45 LEAD
Outcome Attributed to Adverse Event Life Threatening;
Patient Age74 YR
Patient SexFemale
Patient Weight66 KG
-
-