STRYKER ORTHOPAEDICS-MAHWAH EXETER V40 STEM 33MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 0580-1-330
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Pain (1994); Insufficient Information (4580)
Date of Event 02/28/2024
Type of Reportable Event Serious Injury
Event or Problem Description
The stem fracture was found.Revision surgery is not planned yet.
 
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.H3 other text : device not returned.
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture involving an exeter stem was reported.The event was confirmed via clinician review of the provided medical records.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical information by a clinical consultant indicated: "no clinical history or medical information were provided for review.A plain x-ray with significant artifact which is unlabeled and undated shows what appears to be a fracture of an exeter style stem at the midpoint.No significant conclusions can be made as there is no medical history or other information provided.Event confirmation: a stem fracture can be confirmed.Root cause: the root cause of the fracture cannot be ascertained as there is a paucity of information, medical history, and significant artifact over the x-ray." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient is planned to be revised due to fracture of the stem.A review of the provided medical information by a clinical consultant indicated the following: "no clinical history or medical information were provided for review.A plain x-ray with significant artifact which is unlabeled and undated shows what appears to be a fracture of an exeter style stem at the midpoint.No significant conclusions can be made as there is no medical history or other information provided.Event confirmation: a stem fracture can be confirmed.Root cause: the root cause of the fracture cannot be ascertained as there is a paucity of information, medical history, and significant artifact over the x-ray." no further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
The stem fracture was found.Revision surgery is not planned yet.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXETER V40 STEM 33MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key18986187
Report Number0002249697-2024-00506
Device Sequence Number8204311
Product Code JDI
UDI-Device Identifier04546540153227
UDI-Public04546540153227
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173499
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2020
Device Catalogue Number0580-1-330
Device Lot NumberG5694211
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/28/2024
Supplement Date Received by Manufacturer06/24/2024
Initial Report FDA Received Date03/27/2024
Supplement Report FDA Received Date07/15/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/18/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention; Hospitalization;
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-