| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: canada.H3: customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was fractured.No more information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon review of our reporting decision on fracture of articular surface provisionals.A review of all complaints for this product indicates that there has never been an event that has resulted in a death or serious injury.Therefore, zimmer biomet will not be reporting this failure mode going forward.Mechanical (g04) - provisional top.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned device shows signs of repeated use and the locking rail feature on the lateral distal side has cracked / fractured.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|