| Model Number |
24672 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon leak occurred.A 3.0 x 120, 135cm mustang balloon catheter was advanced for dilation.During the procedure, upon first inflation the balloon was found leaking.The device was tested outside the body.It was confirmed that the leak was within balloon material.The procedure was completed with a different device.No complications were reported, and the patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reported address 1:(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon leak occurred.A 3.0 x 120, 135cm mustang balloon catheter was advanced for dilation.During the procedure, upon first inflation the balloon was found leaking.The device was tested outside the body.It was confirmed that the leak was within balloon material.The procedure was completed with a different device.No complications were reported, and the patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reported address 1: (b)(6).Device evaluated by mfr: the mustang was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The rated burst pressure for this device is 24 atmospheres.The returned device was attached to an encore inflation unit.Positive pressure was applied when di water was observed to be leaking from a balloon pinhole located approximately 45mm proximal from the distal markerband.A visual examination of the markerbands identified no issues.A visual and tactile examination identified no kinks or damage to the shaft.A visual and tactile examination identified no damage to the tip.
|
| |
|