BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24672
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon leak occurred.A 3.0 x 120, 135cm mustang balloon catheter was advanced for dilation.During the procedure, upon first inflation the balloon was found leaking.The device was tested outside the body.It was confirmed that the leak was within balloon material.The procedure was completed with a different device.No complications were reported, and the patient was stable.
 
Additional Manufacturer Narrative
E1: initial reported address 1:(b)(6).
 
Event or Problem Description
It was reported that balloon leak occurred.A 3.0 x 120, 135cm mustang balloon catheter was advanced for dilation.During the procedure, upon first inflation the balloon was found leaking.The device was tested outside the body.It was confirmed that the leak was within balloon material.The procedure was completed with a different device.No complications were reported, and the patient was stable.
 
Additional Manufacturer Narrative
E1: initial reported address 1: (b)(6).Device evaluated by mfr: the mustang was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The rated burst pressure for this device is 24 atmospheres.The returned device was attached to an encore inflation unit.Positive pressure was applied when di water was observed to be leaking from a balloon pinhole located approximately 45mm proximal from the distal markerband.A visual examination of the markerbands identified no issues.A visual and tactile examination identified no kinks or damage to the shaft.A visual and tactile examination identified no damage to the tip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18987722
Report Number2124215-2024-18313
Device Sequence Number8265055
Product Code FGE
UDI-Device Identifier08714729793816
UDI-Public08714729793816
Combination Product (Y/N)N
Initial Reporter CountryID
PMA/510(K) Number
K103751
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/15/2026
Device Model Number24672
Device Catalogue Number24672
Device Lot Number0031832181
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/05/2024
Supplement Date Received by Manufacturer08/08/2024
Initial Report FDA Received Date03/27/2024
Supplement Report FDA Received Date09/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/16/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1