• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE "PLASMALOOP", LOOP, MEDIUM, 24 FR., 12°-30°, ESG TCRIS; RESECTION ELECTRODES WITH HF CABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE "PLASMALOOP", LOOP, MEDIUM, 24 FR., 12°-30°, ESG TCRIS; RESECTION ELECTRODES WITH HF CABLE Back to Search Results
Model Number WA47706S
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported, during a therapeutic hysteroscopy/myomectomy procedure using the loop cautery, the loop broke at the connection point to the electrode.The surgeon visualized the uterine cavity the entire time from the break to the retrieval and the loop itself was retrieved, intact, from the uterine cavity.There were no further foreign bodies visualized after the retrieval.The procedure was completed with another device.A fluoroscopy x-ray of the pelvis/uterine cavity was taken, followed by a flat p late x-ray at the conclusion of the procedure.
 
Event or Problem Description
It was reported, the loop broke at the end of the procedure and was removed by trapping the wire with the end of the retractable loop arm without any cautery activated.Another surgeon then proceeded with the patient's breast surgery.There was no delay as the x-ray was done at the end of the completed cases, not in between.This report is being supplemented to provide additional information from the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE "PLASMALOOP", LOOP, MEDIUM, 24 FR., 12°-30°, ESG TCRIS
Common Device Name
RESECTION ELECTRODES WITH HF CABLE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
MDR Report Key18988466
Report Number2429304-2024-00209
Device Sequence Number19575632
Product Code FAS
UDI-Device Identifier14042761085530
UDI-Public14042761085530
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/09/2024
Report Date (Section F) 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA47706S
Device Lot Number1100442429
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date03/14/2024
Event Location Hospital
Date Report to Manufacturer03/14/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/27/2024
Supplement Report FDA Received Date04/09/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient SexFemale
Patient Weight94 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-