Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while using the bd vacutainer® push button blood collection set with pre-attached holder that blood is leaking during blood collection.No patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
D2b: medical device type: jka h.6.Investigation summary: "material #: 367352 lot/batch #: 3292564 bd received 1 sample and 1 photo for investigation.The photo was reviewed and the customer¿s indicated failure mode for leakage was observed.Additionally, the customer sample along with 30 retention samples from bd inventory were evaluated by functional draw testing and the indicated failure mode for leakage with the incident lot was not observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode leakage.Bd determined that the root cause of the indicated failure mode was attributed to the tubing being cut slightly too long to fit securely in the unit blister.The tubing length was shortened which corrected the issue.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends." h3 other text : see h.10.
|
| |
|