Catalog Number C408646
Medical Device Problem Codes Difficult to Remove (1528); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no other similar incidents.The investigation was unable to determine a conclusive cause for the reported noise problem and the difficult to remove.It may be possible that the dragonfly was incorrectly mated to the doc or there was a system issue on the doc causing noise.It also may be possible that the user¿s technique when removing the dragonfly over the guide wire may have contributed to the interaction between the devices, resulting in the reported difficult to remove.However, without having the device to examine, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that noise was heard during the pullback process of the dragonfly catheter.Additionally, the catheter was taken out of the patient and it was wrapped around the unspecified guide wire [resistance removing].The procedure was successfully completed without additional oct.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRAGONFLY¿ OPTIS¿ IMAGING CATHETER
Common Device Name
DIAGNOSTIC INTRAVASCULAR CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT MEDICAL REG#3009600098
4 robbins rd
westford MA 01886
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18990890
Report Number2024168-2024-03767
Device Sequence Number15351250
Product Code DQO
UDI-Device Identifier00183739000654
UDI-Public(01)00183739000654(17)250625(10)9109105
Combination Product (Y/N)N
Initial Reporter CountryLG
PMA/510(K) Number
K141769 
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/25/2025
Device Catalogue NumberC408646
Device Lot Number9109105
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/05/2024
Initial Report FDA Received Date03/28/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-