| Catalog Number |
11012 |
| Medical Device Problem Codes |
Unable to Obtain Readings (1516); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that: "healthcare clinician used a humid-vent pedi filter on a patient in the day clinic and was unable to obtain a co2 reading on the capnograph.Clinical consequences: the filter was replaced and the dr was the able to obtain the desired reading.".
|
| |
|
Event or Problem Description
|
|
It was reported that: "healthcare clinician used a humid-vent pedi filter on a patient in the day clinic and was unable to obtain a co2 reading on the capnograph.Clinical consequences: the filter was replaced and the dr was the able to obtain the desired reading.".
|
| |
|
Additional Manufacturer Narrative
|
|
Qn # (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The sample was returned to the manufacturer for evaluation.The manufacturer reported: "complaint reported that doctor was unable to obtain a co2 reading on the capnograph.Based on the investigation conducted on the returned sample, luer port of the complaint product housing has occluded by the housing material.One actual sample was received.The sample was closely examined by visual inspection.The result of visual inspection is that the luer port is occluded.The mentioned product housing was supplied by our supplier.Assembly process will be conducted in our manufacturing site for the product after they receive the part item from supplier.A device history record review was performed, and no relevant findings were identified.A new non-conformance was raised to further investigate on this issue.Based on the investigation conducted on the returned sample, this complaint is confirmed.The mentioned product housing was supplied by our supplier and is out of the scope from previous supplier corrective action requests and hence, a non-conformance has been issued and root cause analysis and identified corrective actions will be implemented." teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that: "healthcare clinician used a humid-vent pedi filter on a patient in the day clinic and was unable to obtain a co2 reading on the capnograph.Clinical consequences: the filter was replaced and the dr was the able to obtain the desired reading".
|
| |
|